For a structure elucidation process, applying an apodization (or window function) to an NMR spectrum can be a good thing and sometimes can be a bad thing. Making peaks broader can assist in differentiating signals from noise, but can also cause neighboring peaks to mesh into a single resonance. The best approach in handling spectral...

Certain experimental conditions, such as varying the temperature or using different solvents, can influence the interpretation of NMR data. Although it can be useful to collect data with different solvents (e.g. DMSO-d6 or Chloroform-d) or temperatures to decrease peak overlap in a particular region, it can complicate the structure elucidation process by increasing the amount...

When working with multiple NMR datasets, it is key to set a reference point(s) in order to align all the datasets. The elucidation purpose — as opposed to standard NMR purposes of referencing to a standard such as TMS — behind referencing the spectral data to a common point(s) is to facilitate the interpretation and...

Like driving a car, the driver navigates the vehicle within the designated lanes on a roadway, thus using the lane markers as a reference. The same can be said when analyzing spectral data. Correctly referencing multiple NMR data is critical in a structure elucidation process. For the case where 1D and 2D NMR data is...

Among intense NMR resonances, weak resonances can be easily overlooked. This is commonly seen with 1H NMR spectra whereby 1H resonances with complicated couplings are spread out over a larger area in comparison to uncoupled 1H resonances that appear as sharp signals. The 1H NMR spectrum below shows 3 intense resonances at 2.39, 2.66 and...

As with many nitrogen-containing compounds, 15N chemical shift information can be critical in elucidating or confirming a candidate structure. Many elucidators extract 15N chemical shifts from such experiments as 15N NMR, 1H-15N HSQC or 1H-15N HMBC. Below are two tautomers, from a previous blog, that differ in the hybridization state of the nitrogen. The 15N...

The best use of a set of tools for a structure elucidation are those that have a general target or goal in mind. Random experiments for the sake of collecting data are a sure sign of inexperience. With that said, Infrared (IR) spectroscopy is a great tool for identifying or confirming a functional group(s). It...

Like a COSY experiment, an HMBC dataset offers many combinations of atom connectivity. The goal for the elucidator is to assess each correlation and narrow down a set of fragments that support the data. For every correlation in a 1H-13C HMBC spectrum, an elucidator must decide whether a correlation corresponds to a 2J, 3J or...

Nine times out of ten, the consequence of an incorrect assignment of a 2D correlation to a 1D resonance is a structural dead end. That is, the fragments cannot be pieced together to present a clear-cut candidate(s). Dealing with ambiguity in an HMBC spectrum is a common area where the structure elucidation process can go...

The best approach when dealing with severe ambiguity in assigning a correlation(s) to a 1D resonance(s) in a 1H-13C HMBC spectrum is to take into account all possible assignments. The drawback to this approach is it adds to the complexity of the structure elucidation process rather than simplifying it. The HMBC spectrum below illustrates a...